Unusual artifact with mammography film.

Title

Unusual artifact with mammography film.

Creator

Hedrick W R; Poulton T B; Starchman D E; Tobias T E

Publisher

Radiology

Date

1998
1998-03

Description

The authors identified the origin of a high-density artifact with mammography film. An acrylic ramp phantom (thickness, 0-7 cm) was used to create a continuously variable radiation exposure. A perturbation in the curve of optical density versus position along the ramp was observed. Development of the film with the emulsion side down eliminated this artifact.

Subject

*Artifacts; *Mammography/instrumentation; *X-Ray Film; Female; Humans; Imaging; Phantoms; X-Ray Intensifying Screens